Basic Anasazi Proton NMR Tutorial
PNMR Tutorial
Not all the PNMR commands are described here, only the ones you need to generate 1D spectra and save your data to disk.
The PNMR program runs in a window, usually with a blue background. It always displays a listing of instrument configuration settings, which can be changed depending upon the sample under analysis or the type of experiment being run.
Here's what the boot-up window looks like:
The top of the window is used to display the FID as data are acquired. Listed at the bottom of the window are the data acquisition parameters, a brief description of what the parameter does, and the current parameter setting.
It's important to note that the PNMR program issues messages during data acquisition that
appear in the top right corner of this main window, and you should read them before proceeding
with data processing. You need to wait until the PNMR message window disappears after data
acquisition before you switch to the NUTS window for data processing.
PNMR commands are typed in at the command prompt and issued by pressing the 'Enter' key. The command prompt appears at the bottom left of the main PNMR window and looks like this:
H1>
and 'H1>' means that the instrument is ready for a proton analysis. Other prompts indicate that the instrument is ready for other types of analyses. If the prompt is anything other than 'H1>' you should type:
NU H1
and then press the Enter key. This should result in an 'H1>' prompt. Then re-initialize the instrument settings by typing:
RE H1.INI
and then press the Enter key. The instrument should now be ready for operation.
A summary of the most common commands is given below, arranged in the order that you will
most likely use them.
START DATA ACQUISITION
At the command prompt, type:
ZG
and press 'Enter'. This starts data acquisition mode and you can see the FID on screen as the
analysis progresses.
START DATA ACQUISITION AND SET RECEIVER GAIN AUTOMATICALLY
At the command prompt, type:
ZGA
and press 'Enter'. This starts GS mode and automatically selects a suitable Receiver Gain (RG)
setting before switching to data acquisition mode. In data acquisition mode you will see the FID
on screen as the analysis progresses. While the total sample run time is a little longer than using
ZG, it is excellent when analyzing multiple samples of unknown concentrations, such as during
organic laboratory classes. One caveat, however, is that ZGA may have difficulty determining
which of two gain settings is the optimum and oscillate back and forth between them. The gain
value can be seen at the top of the screen while ZGA is running. If oscillation occurs, press Ctrl-K to quit and manually set the Receiver Gain parameter to the lower of the two values, then start
the acquisition using ZG.
CHANGING DATA ACQUISITION PARAMETER VALUES:
To change any of the data acquisition parameters listed above; type the two letter abbreviation, a space, the new value, and then press the 'Enter' key. For example, typing,
RG 10
and then pressing 'Enter' changes the Receiver Gain to 10. Note that when a parameter is
changed, it is updated in the list immediately.
QUITTING THE PNMR PROGRAM:
To quit the program, type:
QUIT
and then press 'Enter'. To quit GS mode or an active data acquisition, press Ctrl-Q to return to
the command prompt, then quit.
WRITE THE CURRENT FID TO A DATA FILE
Newly acquired (raw) data is stored in a temporary file named pnmr.fid which is overwritten with each successive acquisition. To save your raw data with a different file name and/or to a different directory; acquire the data, wait for the command prompt, and then type:
WR <pathname><filename.fid>
and press 'Enter'. Note that there must be a space between WR and <pathname>. An example is:
WR A:\MYDATA.FID
which would write the file MYDATA.FID to a floppy disk in the A: drive.
RUN THE BASIC SHIMMING ROUTINE
Load the sample tube labeled 'water' which should be in the sample rack atop the magnet. Set RG to 5. At the 'H1>' command prompt, type:
SHIM
and press 'Enter'. PNMR enters GS mode and displays a continuous FID and waits for you to set
RG. Press Ctrl-Q to quit GS mode and start the shimming routine. Pull out the drawer under the
right edge of the desk. As the shimming routine runs it will ask you questions about the settings
of the shim controls, and also tell you how to change the settings as it runs. Follow the
instructions on screen. When the optimization is finished, the shimming routine will stop.
Record the new shim settings in the Run Log Book, along with the date, time, your name, where
you're from, and that you performed a Basic Shim.
REINITIALIZE SPECTROMETER HARDWARE
At the command prompt, type:
INI
and press 'Enter'. This should reset the spectrometer and load the default PNMR settings.
GS MODE
GS mode is used in manual mode only during the basic shimming routine. Just remember, to
quit GS mode, press Ctrl-Q or Ctrl-K.
SIze of the file which will hold the collected data. Increasing the size parameter increases the
number of data points collected per second, which can increase the resolution of fine structure in
the final spectrum. SI can be set to a maximum of 32k (32768) but will not affect chemical shift
or integration of the final peaks. SI is linked with SW (see below). Note that more is not
necessarily better, and generally this parameter is left at the default value of 8k (8192).
NS
Number of Scans collected during data acquisition. Generally, acceptable data can be collected
using only one scan, but increasing the number of scans can increase signal-to-noise ratio in the
final spectrum. Note that the NS parameter is displayed as 'X/Y' where X is the number of
Dummy Scans (DS, see below) and Y is the number of data acquisition scans (NS).
DS
The number of Dummy Scans performed before the data acquisition scan sequence. Dummy
scans pulse the sample in exactly the same manner as the data acquisition scans (NS), except that
the receiver is switched off. This allows the sample to respond to the pulse program and relax
before data is acquired. Generally, dummy scans slightly reduce the magnitude of the signal
from the initial data acquisition pulse, but usually improves (narrows) the peak width in the final
spectrum.
RG
Receiver Gain parameter, sets the gain of the receiver coil preamplifier. Setting a higher number
improves sensitivity, but there are limits. RG is dependent upon the concentration of the sample
and thus may require adjustment for each and every different sample preparation. What is most
important in setting RG is to avoid excessive amplification of the signal, which overloads the
preamplifier and distorts the data. When overloading occurs, the FID is displayed in red:
It's usually best to set DS 0 and NS 1 until you find the correct RG setting for your sample, then
set DS and NS to higher values. If you get the error message, reduce RG by 5 and run the sample
again. Continue reducing RG in increments of 5 until the error message no longer appears after
data acquisition. Now you're good to go. However, if you're really fussy, you can increase RG
in increments of 2 until you get the error message again, then back off by two.
The normal exit message displayed if RG does not overload the preamplifier looks like this:
but doesn't really tell you if RG is set too low. If RG is too low, the signal-to-noise ratio of your spectrum will be adversely affected. This can be seen in the NUTS window as a very noisy baseline and relatively small sample peaks in the processed data.
In general, if you aren't familiar with the characteristics of your samples, start data acquisition
using ZGA instead of ZG. Note that RG 20 is the optimum value for the ethylbenzene test
sample, and that RG 5 is optimum for the water sample. See GS Mode for realtime RG
optimization. See ZGA for automatic RG adjustment before data acquisition.
PW
Pulse Width timing parameter. PW sets the length of time that the RF pulse is on, and is
adjusted so that the sample protons respond by aligning their magnetic 'spin' vectors at a specific
angle to the initial magnetic field of the large NMR magnet. Different pulse durations result in
different angles of alignment. A pulse duration resulting in an angle of 90 degrees is called a '90
degree pulse', and results in the strongest signal from the sample. We will only use a 90
degree pulse program in this tutorial, but other timings are also named for their characteristic
alignment angles; '45 degree pulse', etc., and have specific uses. These alternate pulse programs
will be described in more detail in other experiment tutorials.
RD
Relaxation Delay timing parameter. RD sets the length of time that the data acquisition software
waits between successive RF pulsing of the sample. When the sample is RF pulsed it's energy
increases as the protons magnetic 'spin' vectors are forced out of alignment with the initial
magnetic field of the large NMR magnet. When the pulse stops, the sample begins to 'relax' as
the protons magnetic 'spin' vectors return to alignment with the initial magnetic field, but this
relaxation does not occur instantaneously. Depending upon the sample, a short period of time is
required to ensure that all protons in the sample have returned to alignment, and this period of
time is called the Relaxation Delay. If all of the sample protons are not allowed to relax
completely between pulses, phasing of the spectrum and peak integrations will be adversely
effected. See T1 and T2 Experiment Tutorials for more detail on optimization of
RD. While the default time of 1 second is adequate for most small molecule samples, I usually
set RD to 3 seconds to ensure that almost all student sample types will run OK. Note that if
NS=1 then RD has no effect other than increasing your run time.
SW
Sweep Width of the spectrum in Hertz. SW defines the range of frequencies that are swept
during data acquisition. The default setting of 1000 Hz is adequate for almost any sample,
corresponding to a maximum PPM value of 16.67 in the final spectrum. Note that SI and SW are
linked. For a given SI, the wider the sweep width the lower the apparent resolution of the
spectrum. This occurs because a fixed number of data points must be spread across larger and
larger intervals. In almost all cases, however, the default values of SI=8192 and SW=1000 are
quite acceptable.
SF
Spectrometer Frequency. The operating frequency of the local oscillator and the resonance
frequency of protons in the applied 14.092 Gauss magnetic field. The default value is 60.01
MHz. Don't mess with it!
DF
Decoupler Frequency. The operating frequency of the decoupler oscillator. Not used for proton
NMR. The default value is 1 MHz. Don't mess with it!
DP
Decoupler Power. Not used for proton NMR. The default value is 0 dB. Don't mess with it!
T aq.
Total acquisition time for the experiment, given in seconds. This is calculated from the values of
the data acquisition parameters listed above.